How cleanroom validation works
Which parameters are measured, in which occupancy state, and what documents the client receives.

Why validation is needed
Validation confirms that the room meets the specified cleanliness class. Without it there is no certainty that the ventilation and enclosure work as designed.
Occupancy states
Tests are carried out in one of three states: as-built (no equipment), at-rest (equipment installed, no personnel) and operational (work in progress).
What is measured
- airborne particle concentration at sampling points;
- velocity and uniformity of unidirectional flow;
- air change rate;
- pressure differentials between rooms;
- integrity and leak tightness of installed HEPA filters;
- where required — recovery time, temperature and humidity.
Documents
The results are recorded in test reports with measured values and a conclusion on class compliance. They are used at handover and later for periodic retesting.